NextStep Antegrade Chronic Hemodialysis Catheter, REF: AC-15192-SFX, AC-15232-SFX, CS-15192-SFX, CS-15232-SFX…
Title shortened; the full text is below and at the official source.
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-2159-2026
- Category
- Medical devices
- Recall date
- 2026-04-10
- Enforcement report date
- 2026-05-20
- Classification
- Class I
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- ARROW INTERNATIONAL, LLC
From the FDA Enforcement Report dated 2026-05-20. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-2159-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- NextStep Antegrade Chronic Hemodialysis Catheter, REF: AC-15192-SFX, AC-15232-SFX, CS-15192-SFX, CS-15232-SFX, CS-15272-SFX, CS-15312-SFX, CS-15422-SFX, CS-15502-SFX; NextStep Antegrade Hemodialysis Catheter with Arrow Simplicity Micro-Introducer, REF: CS-15192-SFXM, CS-15232-SFXM; NextStep Retrograde Hemodialysis Catheter, REF: AC-15192-X, AC-15232-X, AC-15272-X, CS-15192-X, CS-15232-X, CS-15272-X, CS-15312-X; NextStep Retrograde Hemodialysis Catheter with Arrow Simplicity Micro-Introducer, REF: CS-15192-XM, CS-15232-XM, CS-15272-XM, CS-15312-XM
- Hazard / reason
- 16F dual-valved splittable sheath introducer, the subject of a supplier recall, was included in impacted hemodialysis kits & sets. Sheath introducer may not split as intended, and may result in withdrawal resistance, prolonged procedure time, pain, bleeding, hematoma, tissue injury, vessel wall injury.
- Quantity
- 19,687
- Distribution
- Worldwide distribution: US nationwide and countries of: AE, AL, AR, AT, AU, BE, BG, BS, BR, BW, CA, CH, CL, CN, CO, CZ, DE, EC, EE, ES, FR, GB, GR, GT, HU, ID, IE, IN, IT, JP, KW, KZ, LB, LT, LU, MN, MZ, NA, NL, NO, NP, NZ, PA, PE, PF, PH, PL, PT, QA, RO, RS, RU, SA, SG, SL, SK, SZ, TH, TN, TR, TT, UA, VN, ZA, ZM, and ZW
- Code information
- REF/UDI-DI/Lots: AC-15192-SFX/30801902197758/AC-15192-SFX; AC-15232-SFX/00801902095909/33F23H1018; CS-15192-SFX/30801902195297/33F23E0575, 33F23G0530, 33F23K0187, 33F23K1104, 33F23L0929, 33F24A0275, 33F24C0068, 33F24D1012, 33F25C0020, 33F25E1083, 33F25G0636, 33F25H0292, 33F25H0790, 33F25L0038; CS-15232-SFX/30801902195549/33F23E0576, 33F23G0534, 33F23L0936, 33F24A0279, 33F24C0067, 33F24D1006, 33F24E0684, 33F24H0258, 33F24J0033, 33F24J0085, 33F24J0342, 33F24L0056, 33F24M0231, 33F25A0405, 33F25C0036, 33F25C0664, 33F25D0739, 33F25E0433, 33F25G0716, 33F25H0248; CS-15272-SFX/10801902195217/33F23B0321, 33F23D0859, 33F23H0500, 33F23K0177, 33F23L0943, 33F24A0280, 33F24C0083, 33F24H0042, 33F24H0260, 33F24J0022, 33F24J0038, 33F24J0088, 33F24L0177, 33F25C0028, 33F25E1090, 33F25F0301, 33F25F0514; CS-15312-SFX/30801902197529/33F23G0541, 33F23H0984, 33F23L0951, 33F24A0283; CS-15422-SFX/30801902195327/33F23K0179, 33F23L0956, 33F24A0290, 33F24E0814; CS-15502-SFX/30801902195563/33F23L0959, 33F24A0286, 33F24J0060; CS-15192-SFXM/10801902127157/33F25C0015, 33F25E1084; CS-15232-SFXM/10801902127232/33F25C0671; AC-15192-X/ 00801902095879/33F23F0744, 33F24E0784, 33F25D0685; AC-15232-X/00801902095916/33F23H1013, 33F24J0068, 33F25C0675, 33F25J0589; AC-15272-X/00801902095954/33F23H1023; CS-15192-X/30801902195303/33F23C0886, 33F23E0564, 33F23G0263, 33F23G0533, 33F23K0712, 33F23L0935, 33F24E0187, 33F24E0793, 33F24G0171, 33F24J0017, 33F24J0041, 33F25D0701, 33F25E0476, 33F25F0486, 33F25H0783,…
Shortened. The full text is at the official source.
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class I