Diowave Laser System, REF: Diowave 250W
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-2162-2026
- Category
- Medical devices
- Recall date
- 2026-01-09
- Enforcement report date
- 2026-05-20
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Technological Medical Advancements LLC
From the FDA Enforcement Report dated 2026-05-20. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-2162-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Hazard / reason
- Software update is needed for laser systems, for topical infrared heating to provide temporary relief of minor muscle and joint pain, spasms, stiffness associated with minor arthritis, and to temporarily increase local blood circulation, operating at a maximum average power output of 70W, which is beyond the FDA cleared 60W.
- Quantity
- 16
- Distribution
- US Nationwide distribution in the states of TX, CA, MI, MD, PA, FL, MN, DC, MI.
- Code information
- UDI-DI: G180DIOWAVESTEALTHMAX0. All software versions prior to, but not including V1.81
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II