Philips Azurion. System Code Description (Model Number): Azurion 3 M12 (722221, 722229), Azurion 3 M15 (722222…
Title shortened; the full text is below and at the official source.
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-2170-2026
- Category
- Medical devices
- Recall date
- 2026-03-31
- Enforcement report date
- 2026-05-20
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
From the FDA Enforcement Report dated 2026-05-20. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-2170-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- Philips Azurion. System Code Description (Model Number): Azurion 3 M12 (722221, 722229), Azurion 3 M15 (722222, 722230), Azurion 5 (722281), Azurion 5 M12 (722227, 722231), Azurion 5 M20 (722228, 722232, 722281), Azurion 7 (722282), Azurion 7 B12 (722225, 722235), Azurion 7 B20 (722068, 722226, 722236), Azurion 7 M12 (722078, 722223, 722233), Azurion 7 M20 (722079, 722224, 722234, 722282).
- Hazard / reason
- Potential for the bolts and plastic parts of the Cable Hose Carriers that hold the cable hose to the monitor ceiling suspension may loosen and/or break over time due to the forces applied when the monitor is moved or rotated.
- Quantity
- 1,718 units
- Distribution
- Worldwide distribution - US Nationwide and the country of Canada.
- Code information
- 1. System Code Description: Azurion 3 M12. Model Number [UDI]: 722221 [(01)00884838099203(21)181, (01)00884838099203(21)178, (01)00884838099203(21)93]; 722229 [(01)00884838116726(21)22, (01)00884838116726(21)9, (01)00884838116726(21)16, (01)00884838116726(21)15, (01)00884838116726(21)5, (01)00884838116726(21)3, (01)00884838116726(21)4, (01)00884838116726(21)6, (01)00884838116726(21)2, (01)00884838116726(21)8]. 2. System Code Description: Azurion 3 M15. Model Number [UDI]: 722222 [(01)00884838099210(21)358, (01)00884838099210(21)328, (01)00884838099210(21)389, (01)00884838099210(21)378, (01)00884838099210(21)243, (01)00884838099210(21)392, (01)00884838099210(21)396, (01)00884838099210(21)13, (01)00884838099210(21)379, (01)00884838099210(21)385, (01)00884838099210(21)376, (01)00884838099210(21)366, (01)00884838099210(21)365, (01)00884838099210(21)398, (01)00884838099210(21)274]; 722230 [(01)00884838116733(21)40, (01)00884838116733(21)2, (01)00884838116733(21)41, (01)00884838116733(21)15, (01)00884838116733(21)28, (01)00884838116733(21)38, (01)00884838116733(21)9, (01)00884838116733(21)27, (01)00884838116733(21)13]. 3. System Code Description: Azurion 5. Model Number [UDI]: 722281 [(01)00884838110564(21)401, (01)00884838110564(21)427, (01)00884838110564(21)382, (01)00884838110564(21)425]. 4. System Code Description: Azurion 5 M12. Model Number [UDI]: 722227 [(01)00884838099227(21)265, (01)00884838099227(21)360, (01)00884838099227(21)320, (01)00884838099227(21)338,…
Shortened. The full text is at the official source.
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II