Swan-Ganz Catheter, Models: 131F7/131F7P, 131F7J, 141F7, 151F7, 834F75, 096F6/096F6P, C144F7, 782F75M, 132F5, C146F7…

Title shortened; the full text is below and at the official source.

Agency
U.S. Food and Drug Administration
Recall number
Z-2202-2026
Category
Medical devices
Recall date
2026-04-08
Enforcement report date
2026-05-27
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Becton, Dickinson and Company

From the FDA Enforcement Report dated 2026-05-27. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
Swan-Ganz Catheter, Models: 131F7/131F7P, 131F7J, 141F7, 151F7, 834F75, 096F6/096F6P, C144F7, 782F75M, 132F5, C146F7, 831F75/831F75P, TS105F5, 774F75, 777F8, AIQSGF8;
Hazard / reason
Catheters may, due to a manufacturing process and material changes, leak/break from the (blue) proximal injectate lumen hub, which may lead to infection, medication loss, and/or blood loss.
Quantity
589,223
Distribution
Worldwide - US Nationwide distribution including in the states of TX, SC, CT, GA, VA, MO, NM, UT, NJ, CA, FL, PA, IN, TN, WA, KY, MD, WV, MI, MA, NC, MN, NY, AZ, AL, LA, CO, OK, HI, MS, IL, OH, NE, DC, KS, NV, WI, SD, IA, AR, ID, OR, WY, MT, RI, VT, DE, AK, NH, ND, ME, PR and the countries of Japan, Brazil, Colombia, Australia, New Zealand, Poland, Czech Republic, South Africa, Mauritius, Germany, Israel, Iran, Slovakia, Switzerland, Slovenia, Bulgaria, Cyprus, Saudi Arabia, United Arab Emirates (UAE), Kuwait, Malta, Bahrain, Qatar, Hungary, Morocco, Romania, Lebanon, Tunisia, Oman, Turkey, Austria, Belgium, Luxembourg, France, Netherlands, Italy, Spain, Canary Islands, Portugal, United Kingdom, Northern Ireland, Ireland, Denmark, Sweden, Iceland, Norway, Estonia, Finland, Greece, Chile, Puerto Rico, Canada.
Code information
Model/UDI-DI/Lot(Expiration): 131F7P/00690103000412/65902900(8/1/2026), 65902901(8/3/2026), 65910728(8/2/2026), 65910729(7/29/2026),65910730(7/27/2026), 65910731(7/25/2026), 65910732(7/26/2026), 65910735(7/30/2026), 65910736(8/24/2026), 65910737(7/31/2026), 65916927(8/3/2026), 65916928(8/6/2026), 65916929(8/7/2026), 65916930(8/8/2026), 65916931(8/9/2026), 65916932(8/22/2026), 65916933(8/23/2026), 65944791(8/20/2026),65944792(8/21/2026), 65955857(9/9/2026), 65955858(9/17/2026), 65955859(9/26/2026), 65955860(9/27/2026), 65973739(9/5/2026), 65973740(9/6/2026), 65973741(9/7/2026), 65973742(9/10/2026), 65988571(9/28/2026), 65997478(9/16/2026), 65997479(9/17/2026), 65997481(9/19/2026), 65997482(9/20/2026), 66010504(9/10/2026), 66010505(9/11/2026), 66010506(9/12/2026), 66010507(9/13/2026), 66010508(9/14/2026), 66023074(9/30/2026), 66033088(10/5/2026), 66048930(10/9/2026), 66057629(10/6/2026), 66057630(10/8/2026), 66057944(10/15/2026), 66057945(10/17/2026), 66061552(10/24/2026), 66074999(10/4/2026), 66075318(11/19/2026), 66104892(10/21/2026), 66104893(10/22/2026), 66104894(10/23/2026), 66104895(11/13/2026), 66108277(11/15/2026), 66118638(11/22/2026), 66120310(10/29/2026), 66120312(10/30/2026), 66120313(10/31/2026), 66127027(11/15/2026), 66127028(11/20/2026), 66132570(11/21/2026), 66132571(11/25/2026), 66132572(11/27/2026), 66140447(11/2/2026), 66140448(11/5/2026), 66140449(11/6/2026), 66140450(11/7/2026), 66142269(11/28/2026), 66142274(11/29/2026), 66151008(12/20/2026),…

Shortened. The full text is at the official source.

Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Identifiers

Model numbers
  • 131F7/131F7P
  • 131F7J
  • 141F7
  • 151F7
  • 834F75
  • 096F6/096F6P
  • C144F7
  • 782F75M
  • 132F5
  • C146F7
  • 831F75/831F75P
  • TS105F5
  • 774F75
  • 777F8
  • AIQSGF8

Brands