ARTIS pheno, Model 10849000. interventional fluoroscopic x-ray system

Agency
U.S. Food and Drug Administration
Recall number
Z-2205-2026
Category
Medical devices
Recall date
2026-03-09
Enforcement report date
2026-06-10
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Siemens Medical Solutions USA, Inc

From the FDA Enforcement Report dated 2026-06-10. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Hazard / reason
During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient.
Distribution
US and Worldwide
Code information
UDI: 4056869046877
Recall type
FDA Mandated
Product type
Devices
Classification (FDA wording)