LVP software of the Ivenix Infusion System (IIS); Product Code LVP-SW-0005.

Agency
U.S. Food and Drug Administration
Recall number
Z-2223-2026
Category
Medical devices
Recall date
2026-05-06
Enforcement report date
2026-06-10
Classification
Class I
Status as reported by FDA
Ongoing
Manufacturer / firm
Fresenius Kabi USA, LLC

From the FDA Enforcement Report dated 2026-06-10. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Hazard / reason
Potential software anomaly that can cause the pump to report a false battery-health value (typically 69%) and trigger a battery-depletion alarm that will shut down the pump unless it is reconnected to AC mains power.
Quantity
32 systems
Distribution
US Nationwide distribution in the states of CA, CO, FL, GA, GA, ID, IL, MD, MI, MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI.
Code information
Product Code LVP-SW-0005. UDI-DI: 00811505030122. Software Version 5.10.2
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class I

Identifiers

Barcodes (UPC/GTIN)
  • 811505030122

Brands