LVP software of the Ivenix Infusion System (IIS); Product Code LVP-SW-0005.
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-2223-2026
- Category
- Medical devices
- Recall date
- 2026-05-06
- Enforcement report date
- 2026-06-10
- Classification
- Class I
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Fresenius Kabi USA, LLC
From the FDA Enforcement Report dated 2026-06-10. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-2223-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Hazard / reason
- Potential software anomaly that can cause the pump to report a false battery-health value (typically 69%) and trigger a battery-depletion alarm that will shut down the pump unless it is reconnected to AC mains power.
- Quantity
- 32 systems
- Distribution
- US Nationwide distribution in the states of CA, CO, FL, GA, GA, ID, IL, MD, MI, MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI.
- Code information
- Product Code LVP-SW-0005. UDI-DI: 00811505030122. Software Version 5.10.2
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class I
Identifiers
- Barcodes (UPC/GTIN)
- 811505030122