Ivenix Infusion System (IIS), Large Volume Pump, Model Number: LVP-0004

Agency
U.S. Food and Drug Administration
Recall number
Z-2224-2026
Category
Medical devices
Recall date
2026-05-06
Enforcement report date
2026-06-10
Classification
Class I
Status as reported by FDA
Ongoing
Manufacturer / firm
Fresenius Kabi USA, LLC

From the FDA Enforcement Report dated 2026-06-10. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Hazard / reason
Fresenius Kabi has been made aware via customer complaints that dropping or severe jarring of the Ivenix LVP may cause damage to internal components that may not be apparent or readily visible and the devices therefore remained in service. More specifically, drops or jarring on a hard surface can cause the touchscreen input cable to loosen, resulting in unintended touchscreen behavior on the LVP, including random screen touches and/or lack of screen responsiveness to user input. Under the reported failure condition, unintended touch inputs may register at arbitrary screen locations without user interaction. These inputs occur intermittently and without a predictable pattern.
Quantity
18,444
Distribution
Domestic: CA, CO, FL, GA, ID, IL, MD, MI, MI , MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI, WI;
Code information
Model Number: LVP-0004; UDI-DI: 00811505030320; All Serial/Lot Numbers:
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class I

Identifiers

Barcodes (UPC/GTIN)
  • 811505030320
Model numbers
  • LVP-0004

Brands