GE HealthCare CARESCAPE TELEMETRY SERVER V5 ATO MODEL, REF #2063702-101, Product Code SS9; used in conjunction with…
Title shortened; the full text is below and at the official source.
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-2242-2026
- Category
- Medical devices
- Recall date
- 2026-04-24
- Enforcement report date
- 2026-06-03
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- GE Medical Systems Information Technologies Inc
From the FDA Enforcement Report dated 2026-06-03. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-2242-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- GE HealthCare CARESCAPE TELEMETRY SERVER V5 ATO MODEL, REF #2063702-101, Product Code SS9; used in conjunction with ApexPro CARESCAPE Telemetry Server (CTS) v5.0; Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
- Hazard / reason
- GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Center (Central Station).
- Quantity
- 476 units
- Distribution
- Worldwide distribution.
- Code information
- UDI (01)00840682144773(10)v5.0.2R1C1, Serial Number or Sales Order Number: SS924160026SA, SS921080002SA, SS923080001SA, SS923080006SA, SS922460008SA, SS921320004SA, SS921320010SA, SS921320012SA, SS922120001SA, SS921210005SA, SS921060024SA, SS921060026SA, SS924290077SA, SS923290011SA, SS923290012SA, SS923290013SA, SS923290014SA, SS921520004SA, SS922080001SA, SS922080002SA, SS922120002SA, SS922120003SA, SS924200022SA, SS924450085SA, SS924220065SA, SS921210002SA, SS921210003SA, SS924220066SA, SS924050001SA, SS922170005SA, SS924160027SA, SS924160028SA, SS924160029SA, SS924160031SA, SS924160032SA, SS924160033SA, SS921300003SA, SS921420015SA, SS921420016SA, SS921520002SA, SS920520004SA, SS922250001SA, SS923510002SA, SS923510003SA, SS924160023SA, SS924170017SA, SS924200031SA, SS924200032SA, SS924090011SA, SS924090016SA, SS924090018SA, SS924090020SA, SS924090021SA, SS924090024SA, SS924140003SA, SS924140006SA, SS924140011SA, SS924140012SA, SS924140014SA, SS924170006SA, SS924170007SA, SS924170008SA, SS924170009SA, SS924170010SA, SS924170011SA, SS924170012SA, SS924170013SA, SS924170014SA, SS924170015SA, SS924170016SA, SS921110009SA, SS921420013SA, SS922170004SA, SS922340005SA, SS922290002SA, SS923220003SA, SS923290008SA, SS924020004SA, SS921040002SA, SS921050002SA, SS921060023SA, SS921060025SA, SS921060020SA, SS921060022SA, SS921060016SA, SS921060017SA, SS921060018SA, SS921060021SA, SS921110004SA, SS921110001SA, SS921240007SA, SS920490011SA, SS924430083SA, SS924050023SA,…
Shortened. The full text is at the official source.
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II