PhaseOne Antimicrobial Solution, Model/Catalog Number: 15050
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-2275-2026
- Category
- Medical devices
- Recall date
- 2026-04-27
- Enforcement report date
- 2026-06-10
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Oculus Technologies of Mexico, S.A. de C.V.
From the FDA Enforcement Report dated 2026-06-10. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-2275-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Hazard / reason
- The recall is being conducted due to corrosion of the internal metallic spring within the sprayer component. This condition is caused by material incompatibility between the product solution and the stainless steel spring and may result in visible discoloration and the presence of trace levels of metallic elements (including iron, chromium, and nickel) in the solution during use.
- Quantity
- 5904 units
- Distribution
- US Nationwide distribution in the states of AZ, CA, CO, FL, GA, IL, IN, KY, MI, MS, NC, NJ, NY, TN, TX, UT, VA.
- Code information
- UDI:00860009268616 Lot: 25G263
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Barcodes (UPC/GTIN)
- 860009268616