PhaseOne Antimicrobial Solution, Model/Catalog Number: 15050

Agency
U.S. Food and Drug Administration
Recall number
Z-2275-2026
Category
Medical devices
Recall date
2026-04-27
Enforcement report date
2026-06-10
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Oculus Technologies of Mexico, S.A. de C.V.

From the FDA Enforcement Report dated 2026-06-10. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Hazard / reason
The recall is being conducted due to corrosion of the internal metallic spring within the sprayer component. This condition is caused by material incompatibility between the product solution and the stainless steel spring and may result in visible discoloration and the presence of trace levels of metallic elements (including iron, chromium, and nickel) in the solution during use.
Quantity
5904 units
Distribution
US Nationwide distribution in the states of AZ, CA, CO, FL, GA, IL, IN, KY, MI, MS, NC, NJ, NY, TN, TX, UT, VA.
Code information
UDI:00860009268616 Lot: 25G263
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Identifiers

Barcodes (UPC/GTIN)
  • 860009268616

Brands