Philips Azurion 3M15, Model Numbers: 722064, 722222 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6…
Title shortened; the full text is below and at the official source.
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-2285-2026
- Category
- Medical devices
- Recall date
- 2023-12-07
- Enforcement report date
- 2026-06-10
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
From the FDA Enforcement Report dated 2026-06-10. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-2285-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- Philips Azurion 3M15, Model Numbers: 722064, 722222 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
- Hazard / reason
- During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
- Quantity
- 27 units
- Distribution
- US Nationwide distribution.
- Code information
- UDI (01)00884838085282(21)209, (01)00884838099210(21)100, (01)00884838099210(21)254, (01)00884838099210(21)256, (01)00884838099210(21)257, (01)00884838099210(21)265, (01)00884838099210(21)255, (01)00884838099210(21)99, (01)00884838085282(21)260, (01)00884838085282(21)261, (01)00884838099210(21)69, (01)00884838099210(21)12, (01)00884838085282(21)314, (01)00884838085282(21)259, (01)00884838099210(21)15, (01)00884838099210(21)19, (01)00884838099210(21)23, (01)00884838099210(21)17, (01)00884838085282(21)258, (01)00884838099210(21)55, (01)00884838099210(21)45, (01)00884838085282(21)270, (01)00884838099210(21)187, (01)00884838099210(21)223, (01)00884838099210(21)44, (01)00884838099210(21)113, (01)00884838099210(21)112.
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Possible model numbers (as listed; not verified as model numbers)
- 722064
- 722222