Philips Azurion 7B12/12, Model Numbers: 722067, 722225 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6…
Title shortened; the full text is below and at the official source.
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-2288-2026
- Category
- Medical devices
- Recall date
- 2023-12-07
- Enforcement report date
- 2026-06-10
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
From the FDA Enforcement Report dated 2026-06-10. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-2288-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- Philips Azurion 7B12/12, Model Numbers: 722067, 722225 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
- Hazard / reason
- During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
- Quantity
- 51 units
- Distribution
- US Nationwide distribution.
- Code information
- UDI (01)00884838099265(21)104, (01)00884838085350(21)221, (01)00884838099265(21)153, (01)00884838099265(21)46, (01)00884838099265(21)28, (01)00884838085350(21)250, (01)00884838099265(21)72, (01)00884838099265(21)208, (01)00884838099265(21)41, (01)00884838099265(21)122, (01)00884838099265(21)147, (01)00884838085350(21)225, (01)00884838099265(21)262, (01)00884838099265(21)112, (01)00884838099265(21)110, (01)00884838099265(21)207, (01)00884838099265(21)68, (01)00884838099265(21)177, (01)00884838099265(21)29, (01)00884838099265(21)71, (01)00884838099265(21)37, (01)00884838099265(21)175, (01)00884838099265(21)148, (01)00884838099265(21)205, , (01)00884838099265(21)229, (01)00884838099265(21)228, (01)00884838085350(21)277, (01)00884838085350(21)115, (01)00884838099265(21)157, (01)00884838085350(21)163, (01)00884838099265(21)268, (01)00884838085350(21)75, (01)00884838085350(21)132, (01)00884838085350(21)147, (01)00884838085350(21)208, (01)00884838085350(21)104, (01)00884838085350(21)244, (01)00884838085350(21)234, (01)00884838099265(21)53, (01)00884838099265(21)259, (01)00884838099265(21)55, (01)00884838099265(21)235, (01)00884838099265(21)298, (01)00884838085350(21)70, (01)00884838099265(21)233, (01)00884838099265(21)141, (01)00884838099265(21)289, (01)00884838099265(21)249, (01)00884838099265(21)84, (01)00884838099265(21)111, (01)00884838099265(21)227.
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Possible model numbers (as listed; not verified as model numbers)
- 722067
- 722225