Philips Azurion 7B20/15, Model Numbers: 722068, 722226 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6…
Title shortened; the full text is below and at the official source.
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-2289-2026
- Category
- Medical devices
- Recall date
- 2023-12-07
- Enforcement report date
- 2026-06-10
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
From the FDA Enforcement Report dated 2026-06-10. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-2289-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- Philips Azurion 7B20/15, Model Numbers: 722068, 722226 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
- Hazard / reason
- During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
- Quantity
- 156 units
- Distribution
- US Nationwide distribution.
- Code information
- UDI (01)00884838085367(21)547, (01)00884838099272(21)120, (01)00884838099272(21)122, (01)00884838099272(21)123, (01)00884838099272(21)124, (01)00884838099272(21)125, (01)00884838099272(21)126, (01)00884838099272(21)127, (01)00884838099272(21)153, (01)00884838099272(21)67, (01)00884838099272(21)209, (01)00884838099272(21)271, (01)00884838099272(21)675, (01)00884838099272(21)112, (01)00884838085367(21)467, (01)00884838085367(21)335, (01)00884838099272(21)113, (01)00884838099272(21)121, (01)00884838099272(21)139, (01)00884838099272(21)158, (01)00884838085367(21)455, (01)00884838085367(21)494, (01)00884838099272(21)239, (01)00884838099272(21)105, (01)00884838099272(21)678, (01)00884838099272(21)362, (01)00884838099272(21)114, (01)00884838085367(21)564, (01)00884838099272(21)403, (01)00884838085367(21)418, (01)00884838085367(21)440, (01)00884838099272(21)108, (01)00884838099272(21)545, (01)00884838099272(21)197, (01)00884838099272(21)380, (01)00884838099272(21)281, (01)00884838099272(21)495, (01)00884838099272(21)319, (01)00884838099272(21)339, (01)00884838099272(21)182, (01)00884838099272(21)183, (01)00884838085367(21)64, (01)00884838099272(21)325, (01)00884838099272(21)100, (01)00884838099272(21)272, (01)00884838099272(21)680, (01)00884838085367(21)42, (01)00884838099272(21)255, (01)00884838099272(21)341, (01)00884838099272(21)687, (01)00884838099272(21)350, (01)00884838099272(21)280, (01)00884838085367(21)323, (01)00884838099272(21)270, (01)00884838099272(21)254,…
Shortened. The full text is at the official source.
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Possible model numbers (as listed; not verified as model numbers)
- 722068
- 722226