Avanta Multi-Patient Administration Tube Sets (AVA 500 MPAT (Part 2 of 2)) used with the Medrad¿ Avanta Injection…
Title shortened; the full text is below and at the official source.
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-2296-2026
- Category
- Medical devices
- Recall date
- 2026-04-24
- Enforcement report date
- 2026-06-10
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Bayer Medical Care, Inc.
From the FDA Enforcement Report dated 2026-06-10. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-2296-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- Avanta Multi-Patient Administration Tube Sets (AVA 500 MPAT (Part 2 of 2)) used with the Medrad¿ Avanta Injection System; Catalog number: AVA 500 MPAT (Part 2 of 2); Affected material numbers: 87629007, 60729458, 86566648;
- Hazard / reason
- Supplier changes resulted in a fit issue with the Contrast Flow Control Valve (CFCV) interface on the Avanta Injector System. Multiple customer reports indicate that the stopcock cannot be reliably engaged with the CFCV snap interface. In instances where engagement is achieved through applied force, the connection is unstable, and the stopcock may disengage before or during contrast injection, leading to procedure interruption.
- Quantity
- 11,300 units (1650 US, 9650 OUS)
- Distribution
- Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, DE, FL, IA, IL, IN, KS, KY, LA, MA, MD, ME, MO, NC, NY, OH, OK, PA, SC, TN, TX, VA, WI and the countries of AE, BA, BE, CA, CH, CN, DE, DK, ES, FR, GB, HK, HR, IL, KW, MT, MX, NL, RO, RU, SE, ZA.
- Code information
- Catalog number: AVA 500 MPAT (Part 2 of 2) Affected material numbers: 87629007, 60729458, 86566648; Affected batch numbers: 251702, 252302, 252502, 252602, 252702, 252802, 253304, 253702, 253704, 253804, 254202, 254304, 254704, 254802;
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II