Sharpoint 2.2MM, ANGLED, DOUBLE- BEVEL Model/Catalog Number: 72-2231 Microsurgical Knife
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-2309-2026
- Category
- Medical devices
- Recall date
- 2026-04-23
- Enforcement report date
- 2026-06-10
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- SSC- Surgical Specialties Corporation
From the FDA Enforcement Report dated 2026-06-10. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-2309-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- Sharpoint 2.2MM, ANGLED, DOUBLE- BEVEL Model/Catalog Number: 72-2231 Microsurgical Knife
- Hazard / reason
- 3.0mm surgical knives labeled and packaged as 2.2mm surgical knives.
- Quantity
- 1818 units
- Distribution
- Worldwide - US Nationwide distribution in the states of Texas, Utah, Michigan and the countries of Turkey, Germany, Romania, Poland, United Kingdom.
- Code information
- UDI-DI 00848782027514; Lot Number FA22CHQ
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Barcodes (UPC/GTIN)
- 848782027514