Plum Solo Precision IV Pump, 40001-0401
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-2311-2026
- Category
- Medical devices
- Recall date
- 2026-04-29
- Enforcement report date
- 2026-06-17
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- ICU Medical, Inc.
From the FDA Enforcement Report dated 2026-06-17. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-2311-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Hazard / reason
- Under certain conditions, the pump may not display a soft limit alert, and will not notify the clinician of a soft limit violation before confirming and starting delivery. This may potentially lead to under- or over-delivery.
- Quantity
- 13,613
- Distribution
- Worldwide - US Nationwide distribution in the states of ID, IL, IN, KS, MA, MD, NM, NY, PA, TN, TX and the country of Canada.
- Code information
- UDI-DI M335400011 Plum Solo Precision IV Pump v1.0.3
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II