Allia IGS 3 Pulse angiographic X-ray system
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-2314-2026
- Category
- Medical devices
- Recall date
- 2026-04-20
- Enforcement report date
- 2026-06-17
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- GE Medical Systems, LLC
From the FDA Enforcement Report dated 2026-06-17. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-2314-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Hazard / reason
- GE HealthCare has become aware of a potential issue affecting an electrical cable on certain IGS systems equipped with Gaia generators. This issue could lead to a risk of electrical shock to service personnel performing installation, maintenance, or troubleshooting activities inside the C-FRT cabinet while power is ON.
- Quantity
- 9 units
- Distribution
- Worldwide - US Nationwide distribution.
- Code information
- UDI-DI: 00195278719577; Serial Numbers: M3-24-053, M3-24-124, M3-25-029, M3-25-051, M3-25-053, M3-25-091, M3-25-098, M3-25-124, M3-26-010, M3-26-030
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Barcodes (UPC/GTIN)
- 195278719577