Description/REF: ACCESS TRAY/ASK-04001-SUH-S; ACCESS TRAY/ASK-04001-VM1; STIMUCATH SFTY KIT: 19G CTH, 17Gx4CM…
Title shortened; the full text is below and at the official source.
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-2364-2026
- Category
- Medical devices
- Recall date
- 2026-04-21
- Enforcement report date
- 2026-06-24
- Classification
- Class I
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- ARROW INTERNATIONAL, LLC
From the FDA Enforcement Report dated 2026-06-24. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-2364-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- Description/REF: ACCESS TRAY/ASK-04001-SUH-S; ACCESS TRAY/ASK-04001-VM1; STIMUCATH SFTY KIT: 19G CTH, 17Gx4CM NDL/AB-19604-KS; STIMUCATH SFTY KIT: 19G CTH, 17Gx8CM NDL/AB-19608-KS; STIMUCATH SFTY KIT: 20G CTH, 18Gx8CM NDL/AB-20608-KS; STIMUCATH KIT: 19G CATH, 17G x 9CM NDL/AB-05060-PK; FLEXBLOCK ULTRASOUND KIT: 17Gx9CM NS NDL/ASK-19609-VCU; CONTINUOUS NERVE BLOCK KIT/ASK-19600-CC2; CONTINUOUS NERVE BLOCK KIT 19 GA X 60/ASK-19608-CC; CONTINUOUS NERVE BLOCK KIT 19GA X 60CM/ASK-19608-KP; CONTINUOUS NERVE BLOCK KIT 19GA X 60CM/ASK-19608-SHC; FLEXBLOCK KIT:ECHO CATH, 17Gx8CM STM NDL/FB-19608-K; FLEXBLOCK ULTRASOUND KIT: 17Gx8CM STM ND/FB-19608-KU; FLEXBLOCK KIT:ECHO CATH, 17Gx9CM NS NDL/FB-19609-K; FLEXBLOCK ULTRASOUND KIT: 17Gx9CM NS NDL/FB-19609-KU; CONTINUOUS PERIPHERAL NERVE BLOCK KIT/ASK-19608-US1; EPIDURAL CATHETERIZATION KIT/ASK-05001-SLR1; EPIDURAL CATHETERIZATION KIT/MH-05001; EPIDURAL CATHETERIZATION KIT/WJ-05401; SPINAL ANES/EPIDURAL CATH 19GA X 90CM/ASK-05501-USC1; EPIDURAL NEEDLE KIT/LG-03000; EPIDURAL ANESTHESIA KIT/MTO-03000-VM; EPIDURAL ANESTHESIA KIT/SM-03001; Anesthesia conduction kit/FB-19611-K
- Hazard / reason
- Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.
- Quantity
- 205,685
- Distribution
- US Nationwide distribution including in the states of VA, NY, UT, SC, LA, MN, PA, AZ, CA, MI, OH, FL, CT, KS, IL, TX, WI, GA, NC, MD, OR, AK, NH, IA, ND, OK, NM, AL, MA, KY, WA, AR, CO, IN, ME, TN, WY, NJ, SD, MS, MO, MT, DE, ID, NE, WV, RI, VT, NV, HI, DC, PR.
- Code information
- REF(Material)/UDI-DI/Lot(Batch): ASK-04001-SUH-S/10801902210217, 10801902193275/33F25C0766, 33F25E0112, 33F25G0101, 33F25H0058, 33F25H0725, 33F25J0862, 33F25M0836, 33F25F0160; ASK-04001-VM1/10801902210217, 10801902202106, 10801902193275/33F24L0599, 33F26A0962, 33F24G0280; AB-19604-KS/10801902210217, 10801902209471/33F24L0608, 33F25G0474, 33F25J0814, 33F25M0887, 33F24D0064; AB-19608-KS/10801902210217, 10801902209471/33F24E0312, 33F24F0295, 33F24G0316, 33F24K0082, 33F24L0714, 33F25E0027, 33F25E0904, 33F25H0003, 33F25J0815, 33F25L0337, 33F26B0498, 33F24E0314; AB-20608-KS/)10801902210217/33F25H0004, 33F25J0816, 33F24D0437; AB-05060-PK/10801902209433, 10801902210217/33F24D0002, 33F24E0029, 33F24G0315, 33F24J0484, 33F24K0081, 33F25A0451, 33F25C0747, 33F25E0026, 33F25J0646, 33F24E0028; ASK-19609-VCU/10801902210217, 10801902206876, 10801902214291/33F24E0536, 33F24J0630, 33F24L0605, 33F25C0179, 33F25E0220, 33F25H0140, 33F26B1003, 33F24D0628; ASK-19600-CC2/10801902210057, 10801902210217, 10801902210064/33F24D0393, 33F24E0533, 33F24G0494, 33F24K0251, 33F24L0665, 33F25C0177, 33F25E0217, 33F25H0138, 33F24C0230; ASK-19608-CC/10801902210217, 10801902210064, 10801902202236/33F24G0309, 33F24K0252, 33F24L0639, 33F25B0691, 33F25C0178, 33F25C0761, 33F25E0218, 33F25E1022, 33F25H0750, 33F25J0918, 33F26B0230, 33F24D0394; ASK-19608-KP/10801902210217/33F24G0495, 33F24G0086; ASK-19608-SHC/10801902210217, 40801902202220, 10801902202243/33F24G0496, 33F24J0629, 33F24K0253, 33F24L0573,…
Shortened. The full text is at the official source.
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class I