Advance Aluminum Chloride Gel Clear Hemostatic Gel Kit, Catalog number 504600-28

Agency
U.S. Food and Drug Administration
Recall number
Z-2400-2026
Category
Medical devices
Recall date
2026-05-27
Enforcement report date
2026-06-17
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Inter-Med Llc

From the FDA Enforcement Report dated 2026-06-17. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
Advance Aluminum Chloride Gel Clear Hemostatic Gel Kit, Catalog number 504600-28
Hazard / reason
Some units of Aluminum Chloride Gel contain the package insert for Ferric Sulfate Gel. The incorrect insert misidentifies the product and contains different first aid instructions, most critically in that the Ferric Sulfate insert instructs users to induce vomiting in the case of ingestion, while the correct Aluminum Chloride insert instructs users not to induce vomiting.
Quantity
500
Distribution
US Nationwide distribution in the states of Tennessee and Missouri.
Code information
Model No. 504600-28; UDI-DI 00371347008104; Lot Number 2026-0204
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Identifiers

Barcodes (UPC/GTIN)
  • 371347008104
Possible model numbers (as listed; not verified as model numbers)
  • 504600-28

Brands