Bicarby Dialysate; Model number: RFP-400-G;

Agency
U.S. Food and Drug Administration
Recall number
Z-2401-2026
Category
Medical devices
Recall date
2026-05-07
Enforcement report date
2026-06-17
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Fresenius Medical Care Holdings, Inc.

From the FDA Enforcement Report dated 2026-06-17. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Hazard / reason
There is an error in the labeling of the GTIN on the product. When scanned; however, the case labeling, lot number, and printed labeling on the product correctly identifies the product code and the solution contained in the bag. In the event the incorrect GTIN labeling is used as the singular source for product identification, unrecognized use of a potassium solution with a concentration differing from the prescription may result.
Quantity
4752
Distribution
Domestic: TX.
Code information
Model number: RFP-400-G; UDI-DI: 00840861102839; Lot codes: 25SG06026, 25SG06027;