MOSAIQ Oncology Information System
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-2406-2026
- Category
- Medical devices
- Recall date
- 2026-05-21
- Enforcement report date
- 2026-06-17
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Elekta, Inc.
From the FDA Enforcement Report dated 2026-06-17. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-2406-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Hazard / reason
- Software may fail to load adapted baseline shift plan during treatment delivery when a communication error occurs.
- Quantity
- 12 programs
- Distribution
- Worldwide - US Nationwide distribution in the states of WI, FL, LA, NY, NY, NJ, KS, FL, IA, NM, TX and the countries of Bahrain, Belgium, Canada, China, Denmark, France, Germany, Hong Kong, India, Israel, Italy, Japan, Lithuania, Malta, Mexico, Netherlands, Pakistan, Panama, Poland, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom.
- Code information
- Software Version: MOSAIQ 3.2/UDI: 07340201500071
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Barcodes (UPC/GTIN)
- 7340201500071