Medline convenience kits: INPATIENT CENTRAL LN DRSG KIT DYNDC3525 PICC/CVC/MIDLINE DRSG CHG KIT DT13191A PORT A CATH…

Title shortened; the full text is below and at the official source.

Agency
U.S. Food and Drug Administration
Recall number
Z-2419-2026
Category
Medical devices
Recall date
2026-04-27
Enforcement report date
2026-06-24
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Medline Industries, LP

From the FDA Enforcement Report dated 2026-06-24. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
Medline convenience kits: INPATIENT CENTRAL LN DRSG KIT DYNDC3525 PICC/CVC/MIDLINE DRSG CHG KIT DT13191A PORT A CATH CDS982047S VEIN ABLATION PACK DYNJ62889C
Hazard / reason
Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.
Quantity
49,654 kits total
Distribution
US Nationwide distribution. OUS distribution pending.
Code information
DYNDC3525 UDI-DI 10198459576294 Lot 25KBM985 DT13191A UDI-DI 10653160998914 Lot 26AMA168 CDS982047S UDI-DI 10198459385735 Lot 26AMB202 DYNJ62889C UDI-DI 10198459416453 Lot 26AMA584
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Identifiers

Barcodes (UPC/GTIN)
  • 10198459576294
  • 10653160998914
  • 10198459385735
  • 10198459416453

Brands