Medline convenience kit: KIT: DRESSING CHANGE PED DYNDC3359

Agency
U.S. Food and Drug Administration
Recall number
Z-2426-2026
Category
Medical devices
Recall date
2026-04-27
Enforcement report date
2026-06-24
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Medline Industries, LP

From the FDA Enforcement Report dated 2026-06-24. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
Medline convenience kit: KIT: DRESSING CHANGE PED DYNDC3359
Hazard / reason
Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.
Quantity
49,654 kits total
Distribution
US Nationwide distribution. OUS distribution pending.
Code information
UDI-DI 10195327495589 Lots 25KBK310 25LBN341
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Identifiers

Barcodes (UPC/GTIN)
  • 10195327495589

Brands