Giraffe OmniBed Field Replaceable Unit (FRU), Model Numbers 6600-1056-400 and 6600-1461-500
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-2477-2026
- Category
- Medical devices
- Recall date
- 2026-05-01
- Enforcement report date
- 2026-06-24
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- GE Medical Systems, LLC
From the FDA Enforcement Report dated 2026-06-24. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-2477-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Hazard / reason
- GE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain conditions, for example, impact during transport, one or more of the fasteners that secure the soffit can disengage. If this occurs, a fastener could fall into the patient compartment. In rare circumstances, if multiple fasteners disengage, the canopy soffit can become loose or detach. These conditions could potentially result in patient injury.
- Quantity
- 265 units
- Distribution
- Worldwide - US Nationwide distribution.
- Code information
- Serial Numbers: 5452896 801057486 800977619 801093759 5584569 5587625 5367527 5603110 5371240 5497382
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Possible model numbers (as listed; not verified as model numbers)
- 6600-1056-400
- 6600-1461-500