MEDLINE Medical Procedure Kits labeled as: 1) BASIC HEART PACK, Medline Kit/SKU DYNJ67331B; 2) CARDIAC SURGERY, Medline…

Title shortened; the full text is below and at the official source.

Agency
U.S. Food and Drug Administration
Recall number
Z-2486-2026
Category
Medical devices
Recall date
2026-05-04
Enforcement report date
2026-06-24
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Medline Industries, LP

From the FDA Enforcement Report dated 2026-06-24. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
MEDLINE Medical Procedure Kits labeled as: 1) BASIC HEART PACK, Medline Kit/SKU DYNJ67331B; 2) CARDIAC SURGERY, Medline Kit/SKU DYNJ911589A; 3) CAROTID ENDARTECOMY, Medline Kit/SKU DYNJ36211D; 4) CHLA ECMO / CARDIAC PACK, Medline Kit/SKU DYNJ87973; 5) CHLA ECMO / CARDIAC PACK, Medline Kit/SKU DYNJ87973A; 6) CV PART 1 AND 2, Medline Kit/SKU DYNJ908879; 7) CV PART 1 AND 2, Medline Kit/SKU DYNJ908879C; 8) FEM POP CDS, Medline Kit/SKU CDS983670K; 9) LAP BASIN TRAY, Medline Kit/SKU DYNJ50908B; 10) OPEN HEART LTX CATHETERS KIT, Medline Kit/SKU DYNJ0049032A; 11) OPEN HEART PACK, Medline Kit/SKU DYNJ80549C; 12) OPEN HEART PACK, Medline Kit/SKU DYNJ80549D; 13) RR-ADULT ECMO PACK, Medline Kit/SKU DYNJ69742B; 14) TAVR, Medline Kit/SKU DYNJ906509C.
Hazard / reason
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
Quantity
4801 units
Distribution
Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.
Code information
Medline Kit Number/SKU DYNJ67331B: UDI/DI 10195327013837 (each), 40195327013838 (case), Lot Number 23JBM539; Medline Kit Number/SKU DYNJ67331B: UDI/DI 10195327013837 (each), 40195327013838 (case), Lot Number 23IBF088; Medline Kit Number/SKU DYNJ67331B: UDI/DI 10195327013837 (each), 40195327013838 (case), Lot Number 23FBM203; Medline Kit Number/SKU DYNJ67331B: UDI/DI 10195327013837 (each), 40195327013838 (case), Lot Number 22JBK233; Medline Kit Number/SKU DYNJ67331B: UDI/DI 10195327013837 (each), 40195327013838 (case), Lot Number 22JBB115; Medline Kit Number/SKU DYNJ67331B: UDI/DI 10195327013837 (each), 40195327013838 (case), Lot Number 22IBO304; Medline Kit Number/SKU DYNJ67331B: UDI/DI 10195327013837 (each), 40195327013838 (case), Lot Number 22HBR331; Medline Kit Number/SKU DYNJ67331B: UDI/DI 10195327013837 (each), 40195327013838 (case), Lot Number 22FBM460; Medline Kit Number/SKU DYNJ67331B: UDI/DI 10195327013837 (each), 40195327013838 (case), Lot Number 22DBS093; Medline Kit Number/SKU DYNJ67331B: UDI/DI 10195327013837 (each), 40195327013838 (case), Lot Number 22CBJ072; Medline Kit Number/SKU DYNJ67331B: UDI/DI 10195327013837 (each), 40195327013838 (case), Lot Number 22BBX288; Medline Kit Number/SKU DYNJ67331B: UDI/DI 10195327013837 (each), 40195327013838 (case), Lot Number 21LBU240; Medline Kit Number/SKU DYNJ67331B: UDI/DI 10195327013837 (each), 40195327013838 (case), Lot Number 21JBP452; Medline Kit Number/SKU DYNJ911589A: UDI/DI…

Shortened. The full text is at the official source.

Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Identifiers

Barcodes (UPC/GTIN)
  • 10195327013837
  • 10198459646102
  • 10889942885657
  • 10198459085031
  • 10198459159633
  • 10195327203429
  • 10195327690885
  • 10193489952568

Brands