MEDLINE Medical Procedure Kits labeled as: 1) BNT CV PACK, Medline Kit Number/SKU DYNJ912271; 2) CV PACK PART 1…

Title shortened; the full text is below and at the official source.

Agency
U.S. Food and Drug Administration
Recall number
Z-2488-2026
Category
Medical devices
Recall date
2026-05-04
Enforcement report date
2026-06-24
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Medline Industries, LP

From the FDA Enforcement Report dated 2026-06-24. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
MEDLINE Medical Procedure Kits labeled as: 1) BNT CV PACK, Medline Kit Number/SKU DYNJ912271; 2) CV PACK PART 1, Medline Kit Number/SKU DYNJ31204D; 3) KIT PERIPHERAL VASCULAR MMC, Medline Kit Number/SKU DYNJ905368D; 4) KIT PERIVASCULAR MMC, Medline Kit Number/SKU DYNJ905368B; 5) VP SHUNT PACK, Medline Kit Number/SKU DYNJ46530I.
Hazard / reason
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
Quantity
1248 units
Distribution
Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.
Code information
Medline Kit Number/SKU DYNJ912271: UDI/DI 10198459688324 (each), 40198459688325 (case), Lot Number 26CBQ109; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 25FBO720; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 24KBO387; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 24KBO015; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 24KBD819; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 24KBA736; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 24FBG144; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 24DBO838; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 24BBC453; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 24ABM118; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 23KBS950; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 23IBC670; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 22LBN883; Medline Kit Number/SKU DYNJ31204D: UDI/DI…

Shortened. The full text is at the official source.

Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Identifiers

Barcodes (UPC/GTIN)
  • 10198459688324
  • 10193489957990
  • 10198459698149
  • 10195327047917
  • 10193489338843

Brands