MEDLINE Medical Procedure Kits labeled as: ORAL RESTORATION, Medline Kit Number/SKUDYNJ86253A

Agency
U.S. Food and Drug Administration
Recall number
Z-2490-2026
Category
Medical devices
Recall date
2026-05-04
Enforcement report date
2026-06-24
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Medline Industries, LP

From the FDA Enforcement Report dated 2026-06-24. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
MEDLINE Medical Procedure Kits labeled as: ORAL RESTORATION, Medline Kit Number/SKUDYNJ86253A
Hazard / reason
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
Quantity
0 units
Distribution
Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.
Code information
UDI/DI 10198459556654 (each), 40198459556655 (case). Lot Numbers: 25KBC063, 25JBO955.
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Identifiers

Barcodes (UPC/GTIN)
  • 10198459556654

Brands