MEDLINE Medical Procedure Kits labeled as CLOVIS HYSTEROSCOPY, CSC-N PELVISCOPY, D AND C PACK, D&C HYSTER PACK, FLWR…
Title shortened; the full text is below and at the official source.
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-2505-2026
- Category
- Medical devices
- Recall date
- 2026-05-04
- Enforcement report date
- 2026-06-24
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Medline Industries, LP
From the FDA Enforcement Report dated 2026-06-24. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-2505-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- MEDLINE Medical Procedure Kits labeled as CLOVIS HYSTEROSCOPY, CSC-N PELVISCOPY, D AND C PACK, D&C HYSTER PACK, FLWR GYN PACK, FS LAP TUBAL LIGATION, HILLSDALE D&C / HYSTEROSCOPY, LAVH NEW BRAUNFELS,TVH PACK, etc (see recall documents for a full list of affected products)
- Hazard / reason
- During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
- Quantity
- 37324 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.
- Code information
- Medline Kit Number/SKU DYNJ61063A: UDI/DI 10193489640458 (each), 40193489640459 (case), Lot Number 21KBV571; Medline Kit Number/SKU DYNJ61063B: UDI/DI 10195327193478 (each), 40195327193479 (case), Lot Number 23EBU351; Medline Kit Number/SKU DYNJ61063B: UDI/DI 10195327193478 (each), 40195327193479 (case), Lot Number 23DBQ190; Medline Kit Number/SKU DYNJ61063B: UDI/DI 10195327193478 (each), 40195327193479 (case), Lot Number 23BBE364; Medline Kit Number/SKU DYNJ61063B: UDI/DI 10195327193478 (each), 40195327193479 (case), Lot Number 22IBW094; Medline Kit Number/SKU DYNJ61063C: UDI/DI 10195327451684 (each), 40195327451685 (case), Lot Number 26CMK666; Medline Kit Number/SKU DYNJ61063C: UDI/DI 10195327451684 (each), 40195327451685 (case), Lot Number 26BMG709; Medline Kit Number/SKU DYNJ61063C: UDI/DI 10195327451684 (each), 40195327451685 (case), Lot Number 26AMI644; Medline Kit Number/SKU DYNJ61063C: UDI/DI 10195327451684 (each), 40195327451685 (case), Lot Number 25KMD133; Medline Kit Number/SKU DYNJ61063C: UDI/DI 10195327451684 (each), 40195327451685 (case), Lot Number 25JMA568; Medline Kit Number/SKU DYNJ61063C: UDI/DI 10195327451684 (each), 40195327451685 (case), Lot Number 25IMF717; Medline Kit Number/SKU DYNJ61063C: UDI/DI 10195327451684 (each), 40195327451685 (case), Lot Number 25HMJ595; Medline Kit Number/SKU DYNJ61063C: UDI/DI 10195327451684 (each), 40195327451685 (case), Lot Number 25FMH149; Medline Kit Number/SKU DYNJ61063C: UDI/DI…
Shortened. The full text is at the official source.
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Barcodes (UPC/GTIN)
- 10193489640458
- 10195327193478
- 10195327451684
- 10195327024963
- 10195327165383
- 10193489981728