BD Connecta" BD Luer-Lok" 360 REF 394910 UDI-DI code: 00382903949106 BD Connecta" Stopcocks are used to administer…
Title shortened; the full text is below and at the official source.
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-2514-2026
- Category
- Medical devices
- Recall date
- 2026-05-21
- Enforcement report date
- 2026-06-24
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- BD SWITZERLAND SARL
From the FDA Enforcement Report dated 2026-06-24. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-2514-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- BD Connecta" BD Luer-Lok" 360 REF 394910 UDI-DI code: 00382903949106 BD Connecta" Stopcocks are used to administer fluids or medication through one or two different ports via an IV cannula or extension tube. The ports can be used to sample blood or for hemodynamic monitoring. When used with IV lipid nutritional products, the stopcock device can be used for up to 24 hours.
- Hazard / reason
- Due to stopcock issues when using the plastic luer-lok syringe, the connection at the Port C may over thread or continue to spin rather than tightening securely. This behavior has led some clinical users to perceive the connection as unstable.
- Quantity
- 2,465,500 units
- Distribution
- U.S.: AK, AL, AR, AZ, CA, CO, DC, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, and WV. O.U.S.: Japan
- Code information
- Catalog/Model Number: 394910 UDI-DI code: 00382903949106 Lot Numbers: 4277968 4277973 4282451 4282461 4289801 4290582 4292904 4295848 4299949 4303841 4305108 4306700 5129635 5129642 5129649 5129650 5129651 5139250 5139251 5139252 5150181 5153940 5164061 5164062 5178078 5181442 5183810 5183816 5183824 5183828 5218287 5218291 5218293 5218299 5218302 5241476 5247729 5247731 5251098 5275601 5275602
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Barcodes (UPC/GTIN)
- 382903949106
- Possible model numbers (as listed; not verified as model numbers)
- 394910