Brand Name: Inspire Medical Systems, Inc. Product Name: Inspire Respiratory Sensing Lead Model/Catalog Number: 4340…

Title shortened; the full text is below and at the official source.

Agency
U.S. Food and Drug Administration
Recall number
Z-2515-2026
Category
Medical devices
Recall date
2026-05-11
Enforcement report date
2026-06-24
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Inspire Medical Systems Inc.

From the FDA Enforcement Report dated 2026-06-24. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
Brand Name: Inspire Medical Systems, Inc. Product Name: Inspire Respiratory Sensing Lead Model/Catalog Number: 4340 Software Version: N/A Product Description: The Inspire¿ Respiratory Sensing Lead (Model 4340) is designed to detect respiratory effort. The lead features a pressure sensitive membrane that converts the mechanical energy of respiration into an electrical signal. The lead incorporates a standard connector for coupling to the implantable pulse generator (IPG). Component: N/A
Hazard / reason
Products are labeled with incorrect Use By Date, extending beyond the validated three-year shelf life from date of manufacturer. Material degradation of a product past its Use By Date may result in unplanned or premature revision surgery.
Quantity
50 impacted units
Distribution
Worldwide distribution - US Nationwide and the country of Hong Kong.
Code information
Lot Code: Model No 4340 (USA only lead); UDI-DI [4340] 10855728005946; Inspire part number [4340] 900-014-003; serial numbers T18806, T45012, T45017, T49002, T67098, T67152, T69541, T69542, T69553, T69558, T69562 Model No 4340 (Global lead); UDI-DI [4340] 10810098650301; Inspire part number [4340] 900-014-005; serial numbers T91485, T91486, T91491, T91495, T91513, T91520, T91521, T91527, T91533, T91537, T91540, T91545, T96165, T96168, T96192, T96193, T96195, T96225, T96237, T98118, T98126, T98136, T98137, T98138, T98141, T98143, T98145, T98150, T98153, T98156, T98163, T98165, T98176, T98177, T98178, T98181, T98183, T98190, T96217
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II