Da Vinci ASSY, PSS, SP1098, Part Number: 380601
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-2521-2026
- Category
- Medical devices
- Recall date
- 2026-04-22
- Enforcement report date
- 2026-07-01
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Intuitive Surgical, Inc.
From the FDA Enforcement Report dated 2026-07-01. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-2521-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Hazard / reason
- Operative components of surgical systems potentially distributed with E-brake retainer pin, so in event of a structural column failure, the E-brake pin would prevent E-brake engagement and may result in uncontrolled motion. In addition, if it moves/backs out it may cause restricted column range of motion and/or uncontrolled motion, which may cause patient tissue injury/crushing injury/bleeding.
- Quantity
- 454
- Distribution
- Worldwide - US Nationwide distribution including in the states of CA, GA, OH, PA, NY, MN, TN, OR, IL, MD, NJ, AL, NV, TX, FL, LA, CT, NC, AZ, IA, MO, WA, MI, WV, WI, KY, KS, VA, OK, MA, IN, CO, SC, SD, MS, NH, UT, ME and the countries of Belgium, Denmark, Sweden, France, Japan, Germany, Israel, Switzerland, South Korea, Czech Republic, Austria, United Kingdom, Italy, Spain, Ireland, Saudi Arabia, Greece, Taiwan, Australia, China, Turkey, Poland, United Arab Emirates, Hong Kong, Romania.
- Code information
- UDI: 00886874114605. Serial Numbers: 605408, 606848, 612328, 628399, 609353, 614073, 617794, 616769, 10008062, 635659, 631694, 638771, 641430, 694998, 645433, 643209, 954097, 656203, 764644, 656319, 946098, 667190, 666480, 655562, 664662, 680530, 679266, 683974, 674171, 676962, 682694, 686777, 701248, 696877, 688725, 688225, 697922, 700316, 698277, 702699, 709536, 701751, 717804, 705366, 704731, 724178, 703152, 722803, 693262, 736468, 736338, 740359, 742366, 732233, 696089, 724835, 716520, 709987, 728391, 751529, 731765, 672743, 766366, 711517, 735003, 735252, 746457, 746623, 718940, 721722, 691190, 744094, 750174, 738718, 10211920, 759253, 761448, 752536, 812651, 754935, 790597, 811147, 793019, 761000, 812650, 919988, 919987, 913011, 913012, 926423, 759791, 926422, 938794, 771653, 938795, 766365, 963841, 797872, 769736, 794437, 970568, 970562, 977045, 10003059, 957781, 973824, 804588, 10009706, 936269, 10003509, 930240, 936268, 800142, 10003510, 982782, 967667, 10009707, 973355, 944585, 10065786, 935214, 10020729, 10012963, 10020705, 10031145, 10031144, 973825, 10012962, 10036960, 963856, 10711066, 10035476, 795038, 10043635, 944583, 10072852, 10072853, 966079, 10069899, 10076490, 10087446, 10054478, 10080929, 954096, 10130089, 10165287, 942811, 10180162, 10172589, 10103483, 10103484, 10056518, 10092007, 10246370, 10209961, 10183740, 10232634, 504007, 10231116, 10211880, 10194911, 10245457, 10112470, 10189443, 977044, 10238314, 10644376, 10036961, 10274101, 10112471,…
Shortened. The full text is at the official source.
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Barcodes (UPC/GTIN)
- 886874114605