VOCSN+Pro package, REF: PRT-00853-000; VOCSN+Pro unit, REF: PRT-00490-001
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-2525-2026
- Category
- Medical devices
- Recall date
- 2026-05-08
- Enforcement report date
- 2026-07-01
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Ventec Life Systems, Inc.
From the FDA Enforcement Report dated 2026-07-01. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-2525-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Hazard / reason
- Respiratory system intended to provide continuous/intermittent ventilator support, may have a damaged back-up battery supporting backup alarm function, which may result in the backup alarm not activating during loss of primary power and absence/depletion of backup power sources, which may result in delayed caregiver awareness of a hazardous condition.
- Quantity
- 1
- Distribution
- US Nationwide distribution in the state of UT.
- Code information
- REF/UDI-DI/Serial Number/Lot: PRT-00853-000/00855573007839/155391P/807138, PRT-00490-001/00855573007792/155391/804613
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II