Multi-Snare Set: 5 mm x 125 cm, REF: 147305; 10 mm x 125 cm, REF: 147310
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-2526-2026
- Category
- Medical devices
- Recall date
- 2026-04-21
- Enforcement report date
- 2026-07-01
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- PFM MEDICAL INC.
From the FDA Enforcement Report dated 2026-07-01. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-2526-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Hazard / reason
- A snare catheter including luer lock, indicated for removal or manipulation of a foreign body in the vascular system, failed to meet the biocompatibility requirements for in-vitro-cytotoxicity, which if in contact with a patient's vasculature could result in localized inflammation and tissue irritation.
- Quantity
- 32
- Distribution
- US Nationwide distribution in the states of NY, MD.
- Code information
- REF/UDI-DI/Lot(Expiration): 147305/04042301061829/1049868(10/14/2028); 147310/04042301033567/1049960(10/20/28)
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II