Octopus Evolution AS Tissue Stabilizer, Model TS2500

Agency
U.S. Food and Drug Administration
Recall number
Z-2576-2026
Category
Medical devices
Recall date
2026-05-12
Enforcement report date
2026-07-01
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Medtronic Perfusion Systems

From the FDA Enforcement Report dated 2026-07-01. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Hazard / reason
During the manufacturing process, Medtronic personnel identified an assembly issue where the tubing was incorrectly positioned. The short tube was attached to the bottom of the canister, and the long tube was attached to the top, contrary to the intended design.
Quantity
288 units
Distribution
Worldwide - US Nationwide distribution.
Code information
UDI-DI: 00643169668904, 00613994764379; Serial Numbers: 0232670681 0232709164 0232709168 0232709194 0232709194 0232709199
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Identifiers

Barcodes (UPC/GTIN)
  • 643169668904
  • 613994764379
Model numbers
  • TS2500

Brands