Octopus 4 Tissue Stabilizer, Model 29400

Agency
U.S. Food and Drug Administration
Recall number
Z-2577-2026
Category
Medical devices
Recall date
2026-05-12
Enforcement report date
2026-07-01
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Medtronic Perfusion Systems

From the FDA Enforcement Report dated 2026-07-01. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Hazard / reason
During the manufacturing process, Medtronic personnel identified an assembly issue where the tubing was incorrectly positioned. The short tube was attached to the bottom of the canister, and the long tube was attached to the top, contrary to the intended design.
Quantity
590 units
Distribution
Worldwide - US Nationwide distribution.
Code information
UDI-DI: 00763000543679; Serial Numbers: 0231266463 0231468923 0231478249 0231478253 0231478260 0231651627 0231651634 0231651636 0231651637 0231651638 0231651639 0231651641 0231651642
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Identifiers

Barcodes (UPC/GTIN)
  • 763000543679
Possible model numbers (as listed; not verified as model numbers)
  • 29400

Brands