Octopus 4 Tissue Stabilizer, Model 29400
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-2577-2026
- Category
- Medical devices
- Recall date
- 2026-05-12
- Enforcement report date
- 2026-07-01
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Medtronic Perfusion Systems
From the FDA Enforcement Report dated 2026-07-01. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-2577-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Hazard / reason
- During the manufacturing process, Medtronic personnel identified an assembly issue where the tubing was incorrectly positioned. The short tube was attached to the bottom of the canister, and the long tube was attached to the top, contrary to the intended design.
- Quantity
- 590 units
- Distribution
- Worldwide - US Nationwide distribution.
- Code information
- UDI-DI: 00763000543679; Serial Numbers: 0231266463 0231468923 0231478249 0231478253 0231478260 0231651627 0231651634 0231651636 0231651637 0231651638 0231651639 0231651641 0231651642
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Barcodes (UPC/GTIN)
- 763000543679
- Possible model numbers (as listed; not verified as model numbers)
- 29400