Medline Convenience kits containing recalled Swan-Ganz Catheters CENTRAL AND ARTERIAL LINE TOTE DYNJ8407

Agency
U.S. Food and Drug Administration
Recall number
Z-2608-2026
Category
Medical devices
Recall date
2026-05-28
Enforcement report date
2026-07-08
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Medline Industries, LP

From the FDA Enforcement Report dated 2026-07-08. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
Medline Convenience kits containing recalled Swan-Ganz Catheters CENTRAL AND ARTERIAL LINE TOTE DYNJ8407
Hazard / reason
Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.
Quantity
9,272 kits
Distribution
US Nationwide distribution in the states of AZ, CA, FL, IL, MO, NJ, OH, OK, TN, TX, WI.
Code information
UDI-DI 10198459704543 Lot 26DBB599
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Identifiers

Barcodes (UPC/GTIN)
  • 10198459704543

Brands