Medtronic Harmony Delivery Catheter System

Agency
U.S. Food and Drug Administration
Recall number
Z-2624-2026
Category
Medical devices
Recall date
2026-05-22
Enforcement report date
2026-07-08
Classification
Class I
Status as reported by FDA
Ongoing
Manufacturer / firm
Medtronic Heart Valves Division

From the FDA Enforcement Report dated 2026-07-08. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Hazard / reason
Devices containing inner shaft assemblies from specific manufacturing lots may have an increased potential for distal tip detachment from the delivery system under certain procedural or anatomical conditions due to a manufacturing issue.
Quantity
1881 units
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Canada, Chile, Denmark, Finland, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Japan, Jordan, Mexico, Netherlands, Northern Ireland, Norway, Oman, Poland, Portugal, Puerto Rico, Qatar, Romania, Saudi Arabia, Spain, Sweden, Switzerland, Thailand, United Kingdom.
Code information
GTIN 00763000341367, Lot Numbers: 0012508285, 0012517045, 0012526102, 0012526103, 0012614786, 0012614787, 0012623740, 0012623741, 0012642959, 0012659419, 0012663878, 0012663879, 0012690218, 0012699635, 0012702022, 0012710787, 0012728367, 0012728369, 0012777713, 0012804152, 0012804579, 0012824770, 0012824771, 0012874526, 0012896853, 0012896884, 0012910323, 0012965210, 0012965211, 0012976400, 0012986693, 0013063766, 0013063767, 0013105039, 0013105040, 0013133457, 0013181623, 0013181624, 0013236643, 0013272796, 0013272798. GTIN 00763000544027, Lot Numbers: 0012508285, 0012526103, 0012614786, 0012614787, 0012623740, 0012623741, 0012642959, 0012659419, 0012663879, 0012702022, 0012710787, 0012728367, 0012728369, 0012777713, 0012804152, 0012804579, 0012824771, 0012874526, 0012896853, 0012896884, 0012910323, 0012965211, 0012976400, 0012986693, 0013063767, 0013105039, 0013105040, 0013133457, 0013181623, 0013181624. GTIN 00763000520151, Lot Numbers: 0012575897, 0012642965, 0012728368, 0013272799. GTIN 00763000582951, Lot Numbers: 0012590535, 0012606411, 0012606413, 0012642964, 0012659418, 0012678282, 0012690219, 0012699637, 0012804151, 0012804580, 0012874527, 0012888682, 0012896885, 0012965221, 0012986699, 0013063764, 0013063765, 0013105037, 0013227504. GTIN 00763000918712, Lot Numbers: 0012635392, 0012676200, 0012690207, 0012874528, 0012874529, 0013015539, 0013105038.
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class I

Identifiers

Barcodes (UPC/GTIN)
  • 763000341367
  • 763000544027
  • 763000520151
  • 763000582951
  • 763000918712

Brands