Cyberknife Treatment Deliver System equipped with the Xchange Robotic Collimator Changer; Linear Accelerator

Agency
U.S. Food and Drug Administration
Recall number
Z-2640-2026
Category
Medical devices
Recall date
2026-06-16
Enforcement report date
2026-07-08
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Accuray Incorporated

From the FDA Enforcement Report dated 2026-07-08. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Hazard / reason
In certain cases, a collimator may not fully dock, which triggers a system interlock and prevents motion from the operator console. If personnel then enter the treatment room and initiate motion using the in-room control pendant, the collimator may release and fall from the housing.
Quantity
48 units
Distribution
Worldwide distribution - US Nationwide and the countries of China, Colombia, Czech Republic, France, Hong Kong, Japan, Korea, Republic of, Latvia, Mexico, Myanmar, Saudi Arabia, Taiwan, Turkey, Ukraine.
Code information
All systems prior with software version 11X that utilize the Exchange Table
Recall type
Voluntary: Firm initiated
Product type
Devices