Cyberknife Treatment Deliver System equipped with the Xchange Robotic Collimator Changer; Linear Accelerator
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-2640-2026
- Category
- Medical devices
- Recall date
- 2026-06-16
- Enforcement report date
- 2026-07-08
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Accuray Incorporated
From the FDA Enforcement Report dated 2026-07-08. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-2640-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Hazard / reason
- In certain cases, a collimator may not fully dock, which triggers a system interlock and prevents motion from the operator console. If personnel then enter the treatment room and initiate motion using the in-room control pendant, the collimator may release and fall from the housing.
- Quantity
- 48 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of China, Colombia, Czech Republic, France, Hong Kong, Japan, Korea, Republic of, Latvia, Mexico, Myanmar, Saudi Arabia, Taiwan, Turkey, Ukraine.
- Code information
- All systems prior with software version 11X that utilize the Exchange Table
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II