ABIOMED Introducer Kit 23 Fr, 11 cm, Pump Set Code 1000441, containing Kit, 23Fr Introducer, 11cm, Sterile, Introducer…
Title shortened; the full text is below and at the official source.
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-2684-2026
- Category
- Medical devices
- Recall date
- 2026-06-11
- Enforcement report date
- 2026-07-22
- Classification
- Class I
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Abiomed, Inc.
From the FDA Enforcement Report dated 2026-07-22. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-2684-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- ABIOMED Introducer Kit 23 Fr, 11 cm, Pump Set Code 1000441, containing Kit, 23Fr Introducer, 11cm, Sterile, Introducer Product Code 2000342. heart pump and accessories
- Hazard / reason
- A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.
- Quantity
- 196 units
- Distribution
- Worldwide distribution. US Nationwide, Australia (AU), Austria (AT), Belgium (BE), Brazil (BR), Brunei (BN), Canada (CA), China (CN), Croatia (HR), Czech Republic (CZ), Denmark (DK), Finland (FI), France (FR), Germany (DE), Greece (GR), Hong Kong (HK), Ireland (IE), Israel (IL), Italy (IT), Japan (JP), Kuwait (KW), Luxembourg (LU), Malaysia (MY), Mexico (MX), Netherlands (NL), Norway (NO), Panama (PA), Poland (PL), Saudi Arabia (SA), Serbia (RS), Slovenia (SI), South Korea (KR), Spain (ES), Sweden (SE), Switzerland (CH), Taiwan (TW), Thailand (TH), United Arab Emirates (AE), and United Kingdom (GB).
- Code information
- Pump Set Code 1000441; Introducer Product Code 2000342. GTIN: 00813502012910. All current batches in field.
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class I
Identifiers
- Barcodes (UPC/GTIN)
- 813502012910