VITEK 2 AST-N447 TEST KIT Product reference 424620
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-2715-2026
- Category
- Medical devices
- Recall date
- 2026-06-04
- Enforcement report date
- 2026-07-15
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Biomerieux Inc
From the FDA Enforcement Report dated 2026-07-15. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-2715-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- VITEK 2 AST-N447 TEST KIT Product reference 424620
- Hazard / reason
- Potential for false resistant carbapenem and/or cephalosporin results when analyzing samples with kits from affected lot that have pouch foil fractures.
- Quantity
- 45,900 cards
- Distribution
- International distribution to the countries of Belgium and the Netherlands.
- Code information
- Lot 0853293604 UDI-DI 03573026634551 Product is not registered in the US. OUS only.
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Barcodes (UPC/GTIN)
- 3573026634551