ICU Medical Plum Duo Infusion Pump, v1.2.3: 1) Item Number 400020401 and 2) Item Number 400020403; infusion pump
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-2723-2026
- Category
- Medical devices
- Recall date
- 2026-06-10
- Enforcement report date
- 2026-07-22
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- ICU Medical, Inc.
From the FDA Enforcement Report dated 2026-07-22. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-2723-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- ICU Medical Plum Duo Infusion Pump, v1.2.3: 1) Item Number 400020401 and 2) Item Number 400020403; infusion pump
- Hazard / reason
- In a specific scenario, the pump will not decrement the Volume to be Infused (VTBI) displayed on the Delivery screen and may deliver up to twice the VTBI at the programmed rate.
- Quantity
- 12525 units
- Distribution
- Worldwide distribution - US Nationwide and the country of Canada.
- Code information
- 1) UDI/DI +M335400021, Serial Numbers: 50003704, 50003710, 50004046, 50004072, 50004149, 50004154, 50004156, 50004163, 50004166, 50004168, 50004177, 50004183, 50004205, 50004218, 50004220, 50004249, 50004258, 50004262, 50004263, 50004264, 50004271, 50004300, 50004303, 50004304, 50004305, 50004307, 50004308, 50004310, 50004312, 50004314, 50004316, 50004318, 50004319, 50004323, 50004324, 50004325, 50004326, 50004328, 50004330, 50004331, 50004332, 50004333, 50004334, 50004335, 50004336, 50004338, 50004339, 50004340, 50004341, 50004342, 50004344, 50004345, 50004346, 50004347, 50004348, 50004349, 50004350, 50004351, 50004352, 50004356, 50004358, 50004359, 50004361, 50004363, 50004365, 50004366, 50004370, 50004372, 50004373, 50004374, 50004376, 50004380, 50004382, 50004383, 50004384, 50004385, 50004386, 50004387, 50004389, 50004390, 50004391, 50004392, 50004393, 50004394, 50004395, 50004396, 50004398, 50004399, 50004401, 50004403, 50004405, 50004407, 50004409, 50004410, 50004411, 50004413, 50004415, 50004416, 50004417, 50004424, 50004425, 50004426, 50004428, 50004431, 50004433, 50004435, 50004436, 50004438, 50004439, 50004442, 50004443, 50004444, 50004445, 50004448, 50004449, 50004451, 50004452, 50004455, 50004458, 50004460, 50004461, 50004462, 50004464, 50004465, 50004467, 50004468, 50004469, 50004471, 50004472, 50004473, 50004474, 50004477, 50004481, 50004482, 50004486, 50004487, 50004493, 50004495, 50004499, 50004501, 50004502, 50004506, 50004510, 50004512, 50004513, 50004515,…
Shortened. The full text is at the official source.
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II