Rapid Refill Continuous Injection System, UPN M00566001
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-2725-2026
- Category
- Medical devices
- Recall date
- 2026-06-30
- Enforcement report date
- 2026-07-22
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Boston Scientific Corporation
From the FDA Enforcement Report dated 2026-07-22. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-2725-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Hazard / reason
- Devices contain recalled Medline Namic Angiographic Control Syringe with Rotating Adaptor. There is a risk of syringe loosening or disconnecting from the manifold, resulting in loose connection or complete disconnection. May lead to difficulty connecting syringe, inability to aspirate or inject properly, fluid leakage, or separation of the syringe from the system, potentially requiring device exchange and prolonging the procedure.
- Quantity
- 153,921 (US); 119,274 (OUS)
- Distribution
- US Nationwide distribution. OUS consignee list pending.
- Code information
- UDI-DI (GTIN) 08714729285151 Lots 33786611, 33825949, 33826026, 33949916, 33968571, 33968575, 34079981, 34153104, 34153413, 34153415, 34203075, 34203077, 34153411, 34203079, 34203115, 34386605, 34386609, 34386613, 34429656, 34429658, 34595249, 34693559, 34616613, 34616615, 34712433, 34763427, 34763967, 34805723, 34805725, 34805729, 35006995, 35006999, 35056793, 35208946, 35186845, 35208950, 35397314, 35208952, 35305436, 35498030, 35448316, 35498028, 35551523, 35626939, 35574463, 35688118, 35688460, 35828714,35831703, 35848483, 35877768, 35995282, 36049415, 36159194, 36219882, 36294100, 36303614, 36332230, 31177006, 31180206, 31666278, 31329922, 31356107, 31666276, 31798400, 31798133, 31798135, 31798137, 31852246, 32162450, 32208753, 32255526, 32255528, 32496142, 32661870, 32696046, 32696050, 32696052, 32752697, 32773682, 32791475, 32791477, 32820865, 32838612, 32791479, 33119798, 32977346, 32977348, 33030942, 33120320, 33158008, 33158006, 33354311, 33354315, 33354317, 33403248, 33403252, 33696882, 33696884, 33696886, 33696888, 33750883, 33750885, 33750887, 36332232, 36497262, 36159192, 36587704, 36612263, 36739668, 36785215, 36785217, 36814560, 36845486, 36983883, 36983886, 37291592, 37059471, 37417957, 37417959, 37429808, 37456936, 37472523, 37585745, 37627037, 37654892, 37941477, 37974508, 38033021, 38033023, 38124799, 38143635, 38165214, 38182214, 38201123, 37959187, 38353956, 38374833, 38430201, 33968573, 34386611, 32838610, 38205377, 38429054, 34066207, 34429654,…
Shortened. The full text is at the official source.
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)