Papablic Replacement Brush Heads. Model Number: PAPA317AA.

Agency
U.S. Food and Drug Administration
Recall number
Z-2734-2026
Category
Medical devices
Recall date
2026-05-21
Enforcement report date
2026-07-22
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Papablic Inc

From the FDA Enforcement Report dated 2026-07-22. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Hazard / reason
Potential for brush heads to break.
Quantity
13,400 packs
Distribution
US Nationwide distribution.
Code information
Model Number: PAPA317AA; Amazon FNSKU: X0040USQ0N; Lot Numbers: PS2509A01, PS2511A01, PS2512A01; Date Codes: 2509, 2511, 2512. The recalled product is limited to PAPA317AA replacement brush head packs bearing the above lot numbers/date codes. Products with other lot numbers/date codes are not included in this recall.
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Identifiers

Model numbers
  • PAPA317AA

Brands