Medline Kits containing Zimmer Biomet mixing bowls and spatulas TOTAL KNEE PACK DYNJT6267
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-2736-2026
- Category
- Medical devices
- Recall date
- 2026-06-10
- Enforcement report date
- 2026-07-22
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Medline Industries, LP
From the FDA Enforcement Report dated 2026-07-22. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-2736-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- Medline Kits containing Zimmer Biomet mixing bowls and spatulas TOTAL KNEE PACK DYNJT6267
- Hazard / reason
- Kits contain Zimmer Biomet mixing bowls and spatulas, which were recalled due to seal nonconformities which present risk of sterility breach. Health consequences from use of product include clinically insignificant extension of surgery to find another readily available device.
- Quantity
- 163 kits
- Distribution
- US Nationwide distribution in the states of CA and UT.
- Code information
- UDI-DI 10198459538827 Kit Lots 26DMB644 26CMC467
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Barcodes (UPC/GTIN)
- 10198459538827