Photera 400, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-2742-2026
- Category
- Medical devices
- Recall date
- 2026-03-05
- Enforcement report date
- 2026-07-22
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Daavlin Distributing Company
From the FDA Enforcement Report dated 2026-07-22. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-2742-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Hazard / reason
- Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).
- Quantity
- 2 units
- Distribution
- US Nationwide distribution.
- Code information
- Serial Numbers: 4-01440NB10CL, 2 4-01441NB10CL.
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II