FARADRIVE STEERABLE SHEATH CLEAR, used to facilitate access to the vasculature and into the chambers of the heart…

Title shortened; the full text is below and at the official source.

Agency
U.S. Food and Drug Administration
Recall number
Z-2755-2026
Category
Medical devices
Recall date
2026-06-11
Enforcement report date
2026-07-29
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Boston Scientific Corporation

From the FDA Enforcement Report dated 2026-07-29. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
FARADRIVE STEERABLE SHEATH CLEAR, used to facilitate access to the vasculature and into the chambers of the heart, Material Number 21M402, IFU Reference Number LBL1062
Hazard / reason
Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve damage may result in air ingress, potential air embolism, and associated patient harm. This risk can be mitigated by inserting the dilator straight through the center of the valve and fully advancing it into the sheath body.
Quantity
74,760 units
Distribution
Global Distribution. US Nationwide.
Code information
UDI-DI: 00810087180072; Lots: CL12697, CL12698, CL12701, CL12702, CL12824, CL12832, CL12833, CL12861, CL12888, CL12889, CL12912, CL12913, CL12916, CL12918, CL12968, CL12975, CL12977, CL13003, CL13016, CL13019, CL13045, CL13047, CL13059, CL13078, CL13079, CL13101, CL13102, CL13120, CL13121, CL13140, CL13141, CL13158, CL13159, CL13195, CL13211, CL13218, CL13232, CL13233, CL13259, CL13290, CL13339, CL13367, CL13400, CL13421, CL13516, CL13518, CL13572, CL13577, CL13615, CL13645, CL13669, CL13707, CL13720, CL13733, CL13821, CL13834, CL13857, CL13926, CL13941, CL13976, CL13989, CL14026, CL14052, CL14078, CL14121, CL14138, CL14170, CL14186, CL14225, CL14248, CL14285, CL14344, CL14352, CL14374, CL14395, CL14438, CL14487, CL14499, CL14517, CL14580, CL14632, CL14669, CL14690, CL14710, CL14748, CL14768, CL14967, CL15004, CL15008, CL15066, CL15106, CL15127, CL15190, CL15238, CL15271, CL15317, CL15357, CL15396, CL15426, CL15473, CL15489, CL15567, CL15636, CL15639, CL15753, CL15774, CL15807
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Identifiers

Barcodes (UPC/GTIN)
  • 810087180072

Brands