Hotline Fluid Warming Set, single use device, is designed to warm blood and intravenous fluids and deliver them to the…
Title shortened; the full text is below and at the official source.
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-2759-2026
- Category
- Medical devices
- Recall date
- 2026-06-24
- Enforcement report date
- 2026-07-29
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- ICU Medical Inc.
From the FDA Enforcement Report dated 2026-07-29. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-2759-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- Hotline Fluid Warming Set, single use device, is designed to warm blood and intravenous fluids and deliver them to the patient's intravenous access site at normothermic temperatures under gravity flow conditions, List Number L-70
- Hazard / reason
- ICU Medical became aware of an increased likelihood of leakage in specific Lots of L-70 HOTLINE Fluid Warming Sets due to potentially inadequate solvent application at the tubing to return connector bond.
- Quantity
- 39,750 units
- Distribution
- Worldwide - US Nationwide distribution.
- Code information
- UDI-DI: 30695085407007; Lot Numbers: 4293568, 4293569, 4297510, 4297509, 4326087, 4337724, 4337725, 4346349
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Barcodes (UPC/GTIN)
- 30695085407007