Ventriculostomy Kit, REF: 46154

Agency
U.S. Food and Drug Administration
Recall number
Z-2768-2026
Category
Medical devices
Recall date
2026-06-17
Enforcement report date
2026-07-29
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Medtronic Neurosurgery

From the FDA Enforcement Report dated 2026-07-29. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Hazard / reason
Ventriculostomy devices, used during ventriculostomy procedures to gain access to the ventricular system, may exhibit endotoxin levels that exceed the acceptable limit for devices that come into contact with cerebrospinal fluid; endotoxin exposure can cause an acute inflammatory process.
Quantity
167
Distribution
Worldwide - US Nationwide distribution in the states of UT, AL, CA, NC, GA, AR, LA, KY, LA, KY, IL, OK and the country of Costa Rica.
Code information
UDI-DI: 00673978947396, Lot: 2025-08779
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Identifiers

Barcodes (UPC/GTIN)
  • 673978947396

Brands