Permobil M3 Corpus Power Wheelchair with Heavy Duty Shock REF:120031, Model version 3

Agency
U.S. Food and Drug Administration
Recall number
Z-2795-2026
Category
Medical devices
Recall date
2024-02-29
Enforcement report date
2026-08-05
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Permobil Inc

From the FDA Enforcement Report dated 2026-08-05. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
Permobil M3 Corpus Power Wheelchair with Heavy Duty Shock REF:120031, Model version 3
Hazard / reason
This recall has been reported retroactively. The power wheelchair firmware lacks a speed restriction to maintain stability when the seat lift exceeds 50 mm, which may cause the user to fall from the wheelchair or the wheelchair to tip over,
Quantity
140 units
Distribution
Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KY, ME, MI, MN, MO, MS, MT, NC, ND, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV, WY and the country of CAN.
Code information
All Units operating with firmware version 1.26.2/UDI:(01)17330818345674
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Brands