Halyard convenience kits containing recalled Medline Namic Angiographic Star Off Handle Manifolds. HEART CATH, SELF…
Title shortened; the full text is below and at the official source.
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-2798-2026
- Category
- Medical devices
- Recall date
- 2026-04-30
- Enforcement report date
- 2026-08-12
- Classification
- Class I
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- AVID Medical, Inc.
From the FDA Enforcement Report dated 2026-08-12. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-2798-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- Halyard convenience kits containing recalled Medline Namic Angiographic Star Off Handle Manifolds. HEART CATH, SELF REGIONAL SELF131-05
- Hazard / reason
- Affected kits contain Medline Namic Angiographic Star Off Handle Manifolds, which were recalled by the manufacturer due to presence of particulates within fluid path of the Manifolds. Particulate within the fluid path has the potential to be introduced into blood circulation and become lodged within blood vessels.
- Quantity
- 5168 kits
- Distribution
- US Nationwide distribution in the states of NC and TX.
- Code information
- UDI-DI 10809160408650 Lots 1618351 1621441 1627930 1635908 1644501 1646021 1651957 1654887 1656581 1656582 1659713 1670884 1670885 1671950
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class I
Identifiers
- Barcodes (UPC/GTIN)
- 10809160408650