L-Cath Midline S/L Catheter 20ga (3F) 1mm x 20cm, REF: 384620
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-2804-2026
- Category
- Medical devices
- Recall date
- 2026-06-15
- Enforcement report date
- 2026-08-05
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Argon Medical Devices, Inc
From the FDA Enforcement Report dated 2026-08-05. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-2804-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Hazard / reason
- Catheters, indicated for short/long-term peripheral access to central venous system for the administration of fluids, medications, nutrients; and sampling, may experience leaking and/or cracking at the hub, which may lead to replacement, inability to flush/aspirate catheter, catheter occlusion, therapy interruption, blood loss, tissue irritation, bloodstream infection.
- Quantity
- 20
- Distribution
- US: AR, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, MA, MD, MI, MN, MO, MS, NC, ND, NJ, NV, NY, OR, TN, TX, UT, VA, WA. OUS: MYS, CAN, SAU
- Code information
- REF/UDI-DI/Lot(Expiration): 384620/00886333210008/11490543(7/28/2026)
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II